Standard Thermocouple
MODEL : MTC - 02
Type : R, S
Temperature Scale: ITS-90
Sheath Material: K-99 grade ceramic
- Introduction
- Specifications
- Key Features
Product Details
Reference standard Type ‘R’ & ‘S’ thermocouples are made from High purity reference grade platinum and platinum rhodium alloy. It covers Temperature range up to 1500°C. It is commonly used as reference standard to calibrate Industrial thermocouples.
This sensors are assembled with 6mm OD, 450mm or 600mm long gas tight 99.7% high Alumina sheath. This type thermocouples, besides the low ageing drift, also offers the advantage of a low basic tolerance value.
This model with Cold Junction compensation probe provides much accurate temperature measurement possibilities. The Cold Junction probe is inserted in to ICE water mixture. This is necessary for the Precision measurement.
| Model | MTC - 02 |
|---|---|
| Range |
(a) 200 to 1500°C (b) 100 to 1200°C |
| Type | R, S |
| Accuracy | Class 1 (0.6°C or 0.1% of Temperature whichever is greater) |
| Temperature Scale | ITS-90 |
| Sheath Material | K-99 grade ceramic |
| Thermocouple Wire Dia | 0.5mm |
| Probe Dia & Length | 6mm OD & 450mm Length |
| Reference junction Sheath Dimensions | 250mm Long, 5mm Diameter |
| Cable Length between sensor and CJC sensor | 700mm |
| Cable Length from CJC sensor | 1.5 meter Length |
| Handle Dimension | 20mm OD & 100mm Length |
| Calibration Certified | ISO/IEC 17025 Accredited 5 point |
Note: Due to continuous product improvements, Published specifications may change without notice.
High Stability
Long Life
Low Drift
Wide temperature range